Scan post-process step blocked out with EM "Complete the following scans before continuing to the Analyze page" using TS 1.18.9.2
23/12/2025General Description
WRN message appears: "Complete the following scans before continuing to the Analyze page" on Unite 1.8.13.3 when importing streams using TRIOS Scanning Module 1.18.9.2.
This blocks the next step (Analyze) required for post-processing and sending the case to a lab.

Although the Send button is enabled, attempting to send the case results in the following error message: “An issue occurred, the case does not have any scans that are ready to send.”
Environment
- Multi-device / Production
Preconditions
- Recently updated to the latest TRIOS 1.18.9.2 and Unite 1.8.13.3 module versions.
Steps to Reproduce
- Create a new case using Restorations and enable Pre-preparation, Emergence Profile, or both.
- Import streams using Support Mode.
- Pre-preparation scan, antagonist scan, and bite scan import successfully.
-
When importing the regular surface scan (upper scan or lower s
can), the Analyze page step becomes greyed out, blocking the workflow. - Warning message appears:"Complete the following scans before continuing to the Analyze page."
- Despite the Send button being enabled, sending the case triggers the error message:"An issue occurred, the case does not have any scans that are ready to send." when trying to send the case, the scan can not be post-processed. **
_______________________________________________________________________________________________________________________________________________________________________
Displayed on exception log as:
[ERR] (Source: ThreeShape.SendCommunicate) InvalidOperationException: The case does not have any scans that are ready to send at a.R.aRY.N(ThreeShape.CommonDataAccess.Entity.ICase);ThreeShape.SendCommunicate.dll;100663500;64 at a.R.aRY+j.MoveNext();ThreeShape.SendCommunicate.dll;100663560;84 at System.Runtime.ExceptionServices.ExceptionDispatchInfo.Throw();System.Private.CoreLib.dll;100681135;17 at System.Runtime.CompilerServices.TaskAwaiter.HandleNonSuccessAndDebuggerNotification(System.Threading.Tasks.Task, System.Threading.Tasks.ConfigureAwaitOptions);System.Private.CoreLib.dll;100689381;45 at ThreeShape.Dental.DentalDesktop.Client.Internal.AppServices.AppServicesWorkflow+Q.MoveNext();DDT_3Shape.DentalDesktop.Client.Internal.dll;100669013;475 at System.Runtime.ExceptionServices.ExceptionDispatchInfo.Throw();System.Private.CoreLib.dll;100681135;17 at ThreeShape.Dental.DentalDesktop.Client.Internal.AppServices.AppServicesWorkflow+Q.MoveNext();DDT_3Shape.DentalDesktop.Client.Internal.dll;100669013;1073 at System.Runtime.ExceptionServices.ExceptionDispatchInfo.Throw();System.Private.CoreLib.dll;100681135;17 at System.Runtime.CompilerServices.TaskAwaiter.HandleNonSuccessAndDebuggerNotification(System.Threading.Tasks.Task, System.Threading.Tasks.ConfigureAwaitOptions);System.Private.CoreLib.dll;100689381;45 at a.R.aRY+g.MoveNext();ThreeShape.SendCommunicate.dll;100663548;277 2025-12-23 12:25:15.090 -05:00 (36660:0001) [INF] MessageBox: An issue occurred
Workaround!
Export DCM files using support mode and importing those on a new case, the case should be able to be sent but unfortunately, to do perform any changes on the surface scans.
BUG-related
This issue is currently under investigation on Jira ticket: [BM-8094] [ SR:HIGH ] Scan post-process step blocked out with EM "Complete the following scans before continuing to the Analyze page" using TS 1.18.9.2 - Jira please link your CRM label cases