Filip

Filip Poswiata

1

Article
Last Month

100.0%

How to customize the report types in Ortho Analyzer

Published November 7th, 2023 by Filip Poswiata

In the Ortho Control Panel you can customize the report types according to their specifications. You can create as many different report types as needed. For a step-by-step video click here. Step 1:  Open Ortho Control Panel on the desktop and go to the Custom Report Configuration section and select Report types.  Step 2: To create a new report type

4824 Views 1 min

Export and Save icons are missing on the Toolbar in Ortho Analyzer

Published November 7th, 2023 by Filip Poswiata

Error Overview "Export" and "Save" icons are parts of File Management group of icons in Ortho Analyzer. If File Management is disabled, both icons will be missing on the Toolbar. Resolution Steps Step 1: Right click on the Toolbar and enable File Management. Step 2: "Save" and "Export" icons will appear on the Toolbar.

4521 Views 1 min

How to download and update libraries and templates in Ortho System

Published November 7th, 2023 by Filip Poswiata

Open Orthodontics System Control Panel Click on Download Center Click on “Refresh” Select the libraries you wish to download and click “download” To make an update on a library whenever “New version is available” select the library and click “download” The library should now say “up-to-date” This means the library is downloaded and running the lates

9538 Views 1 min

Part of scan disappears after trimming in Dental Designer or Model Builder

Published November 8th, 2023 by Filip Poswiata

Difficulties with trimming scans in Dental Designer With some of the scans it may be difficult to trim in Dental Designer or Model Builder - a big part of scan is trimmed away. The issue usually issue happens with 3rd party scans, but can also occur with some of the TRIOS scans. Affected DS versions: mostly DS 2018 and 2019. For Model Builder it's p

7698 Views 1 min

Site FTP Inbox settings in Dental System Control Panel are inactive (cannot be modified)

Published November 8th, 2023 by Filip Poswiata

Site FTP Inbox settings in Dental System Control Panel are inactive (cannot be modified) FTP Inbox settings for any lab/manufacturer site in Dental System Control Panel are grayed out and cannot be modified. Root cause: encryption on the dongle prevents from modifying FTP site settings. Solution: please contact support@3shape.com to verify if the en

5175 Views 1 min

How to export a patient file and import it on a second computer that has the software installed

Published November 7th, 2023 by Filip Poswiata

Transfer patient data between computers If one wants to have a specific patient on another computer that has the software installed and is not connected through a client/server setup, what can be done is export the patient's fie and import it on the second system following the next steps:  Step 1: Go to the patient's list  Step 2: Select the patient

6198 Views 1 min

"Incomplete or invalid order. Missing or invalid manufacturer for digital models" when receiving Trios study model orders in 3Shape Communicate Inbox

Published November 8th, 2023 by Filip Poswiata

"Incomplete or invalid order. Missing or invalid manufacturer for digital models" when receiving Trios study model orders in 3Shape Communicate Inbox On some systems when receiving a Trios study model order (a scan with no indications selected), the order may show as empty  in the Inbox and have the following error message: "Incomplete or invalid or

5566 Views 1 min

What licenses are required to import, design restorations and digital models on 3rd party scans in Dental System

Published November 8th, 2023 by Filip Poswiata

Licensing Requirements for Importing 3rd Party Scans in Various Dental System Versions In order to be able to import 3rd party scans into Dental System, design restorations in Dental Designer and create digital models in Model Builder, certain licenses will be required. There have been some changes in the requirements throughout different versions o

9838 Views 1 min

How to design a retainer in Appliance Designer

Published November 7th, 2023 by Filip Poswiata

You can design a retainer in Appliance Designer within the Ortho System by following the steps outlined below. Create/choose a model set. Create a New Appliance. Choose the upper or lower jaw, right-click on it, and Create “Bar”. Set any component properties if needed in the pop-up window and click on “OK”. Perform “Setup insertion direction”. Defin

8438 Views 1 min

Error Message: SOP class (91.2.840.10008.5.1.4.1.1.13.1.2) not supported when importing DICOM file in Ortho Analyzer

Published November 7th, 2023 by Filip Poswiata

If a DICOM file cannot be imported into Ortho Analyzer and the error “Error: SOP class (91.2.840.10008.5.1.4.1.1.13.1.2) not supported” appears, it means the SOP class used to generate the DICOM file is not supported for import. To resolve this issue, the DICOM must be generated according to the parameters specified in the DICOM Conformance Statemen

5532 Views 1 min

"Abutment top cap angle violated" on Finalize step of abutment design

Published November 8th, 2023 by Filip Poswiata

"Abutment top cap angle violated" on Finalize step of abutment design "Abutment top cap angle violated. Angle between implant direction and abutment top cap direction is more than the maximum limitation" error message popping up on Finalize step of abutment design. Affected DS versions: mostly in DS 2018 and later, due to updated FDA regulations. Ro

7536 Views 1 min

Invalid Boundary (Boundary.Count=2) on the Virtual Base Creation step

Published November 7th, 2023 by Filip Poswiata

Handling “holes or artifacts” errors in scans This error appears if scans contain holes or artifacts within the spline area (i.e. on teeth) Solution: Try to correct the spline avoiding holes if it's possible or use any third-party software to close holes and remove artifacts.

6317 Views 1 min

"Unable to import the order. Unsupported material file version" when importing order into Dental Manager

Published November 8th, 2023 by Filip Poswiata

Error with importing an order When importing an order into Dental Manager there is an error message: "Unable to import the order [order file location]. Unsupported material file version: [Dental System version]"  Affected Dental System versions: any version. Root cause: the user is trying to import an order which was created in a higher version of D

5699 Views 1 min

Virtual Setup Animation: A Quick Guide

Published November 7th, 2023 by Filip Poswiata

The Virtual setup Animation is a feature that generates an animated file from the treatment planning, virtual setups. It includes extractions, teeth movements and Inter-proximal reduction (IPR).  How to use Step 1: Open the case.  Step 2: Go to Virtual setup tab. Step 3: Complete the treatment planning. Step 4: To generate the animation click the gr

5910 Views 1 min

"No more saves available in dongle" on Save step in DentalDesigner even if the user has CAD points

Published November 8th, 2023 by Filip Poswiata

Issues with saving designs When trying to save a design there may be an error message: "No more saves available in dongle for this type of order. Please contact your distributor if you want to change this". Error message appears even though the user has enough CAD points on the dongle. Affected DS version: any version. Root cause: The customer has b

5886 Views 1 min

"Failed to morph anatomy shape to its die interface" on Anatomy design step in Dental Designer

Published November 8th, 2023 by Filip Poswiata

Issue with morphing anatomy shape  In some design cases there may be an error message "Failed to morph anatomy shape to its die interface. Library tooth model is placed far away from the margin line. Please place it closer to the margin line" when going to the Anatomy design step in Dental Designer. Most commonly the issue is reproducible on cases w

4923 Views 1 min

Firewall Port Requirements for Dental System Installation and Use

Published November 8th, 2023 by Filip Poswiata

The list of ports was updated for DS 2025, and some of the components (programs\services) may not be present in the lower versions. If either of the ports is not open, the user should contact their local IT assistance to open the ports.   Dental System Port Protocol Program   /   Service 27027 TCP C:\Program   Files   (x86)\3Shape\Dongle   ServerSer

17435 Views 1 min

What is required for integrating Lab Management System with 3Shape Dental System

Published November 8th, 2023 by Filip Poswiata

Integrating Lab Management Systems with 3Shape Dental System 2020 Starting with 3Shape Dental System 2020 there is an integration with global Lab Management Systems (LMS) for efficiency in labs using both systems. The integration synchronizes case information and status across Dental System and LMS via 3Shape Communicate. This eliminates the need fo

8004 Views 1 min

How to change the patient folder label of Ortho Data

Published November 7th, 2023 by Filip Poswiata

Customizing the folder labels for the patients in ortho data can be done by customizing in the Ortho Control Panel. In order to customize the label generation: Open Ortho Control Panel Go to system settings Scroll down to where it says define template for Patient ID generation Proceed to fill in the Patient-ID template fields according to the prefer

5672 Views 1 min

"Inlay UNN - Die interface model could not be produced" when try to create margin line on die interface step

Published November 8th, 2023 by Filip Poswiata

"Inlay UNN - Die interface model could not be produced" when try to create margin line on die interface step When try to create margin line for crowns or inlays on die interface step EM: "Inlay # - Die interface model could not be produced. Unable to define the margin line using the die interface settings. Please try to change the parameters of die

5057 Views 1 min

"Socket Error #11001 Host not found" when opening Dental Manager or Dental System Control Panel

Published November 8th, 2023 by Filip Poswiata

Missing subscription  When opening Dental Manager or Dental System Control Panel there is an error: "No subscription for this application the dongle is either missing or there is no more application subscription available in the dongle. Please set the 3Shape Dongle Service under System Settings. [Socket Error #11001 Host not found]": Affected Dental

6840 Views 1 min

How to download an order manually using 3Shape Communicate Advanced Download tool

Published November 8th, 2023 by Filip Poswiata

Manual case download in 3Shape Communicate for Dental System If for some reason it is not possible to receive or accept some of the cases in 3Shape Communicate Inbox in DS-2018 or later, there is a possibility to download them manually using Advanced Download tool.  Procedure of manual case download: Log into your account on 3Shape Communicate porta

7747 Views 1 min

Error Message: Unable to create shell. Try to change spline. During the shell workflow in appliance designer

Published November 7th, 2023 by Filip Poswiata

This error is displayed when focal points are not correctly placed. At least two focal points must be set on each tooth to continue creating the shell accurately. Resolution In order to resolve the issue, at least two points must be located on each tooth for the shell was created correctly, once they are placed correctly the shell ca be created.

5515 Views 1 min

Holes in scans when redesign the order with digital impression scans

Published November 8th, 2023 by Filip Poswiata

Holes are generated when redesigning the order with digital impression scans When redesign the order with digital impression scans, holes can be generated: Root cause: Prepared scans are used for redesign. The issue is fixed for DS 20.1.1. The issue can happen, when the scans were imported manually as prepared. In this case raw scans will be replace

4478 Views 1 min

How to disable auto-segmentation in Ortho System 2021.1

Published November 7th, 2023 by Filip Poswiata

Auto-segmentation allows the preparation of teeth automatically with high-quality results and saves production time by activating a wide range of features to assist with aligner production. To deactivate using the auto-segmentation feature in the Ortho Control Panel:  Open Ortho System 1.10.0.7. Proceed to Ortho Control Panel -> Settings. Select

5502 Views 1 min

"CAM output could not be generated. Manufacturing process method is not defined for the order" when generating CAM output in Dental Manager

Published November 8th, 2023 by Filip Poswiata

Error when generating CAM output in Dental Manager When trying to generate CAM output in Dental Manager, there may be an error message “CAM output could not be generated. Manufacturing process method is not defined for the order [order_ID]. You can set the manufacturing process method in Control Panel > manufacturing process” Affected Dental Syst

5974 Views 1 min

How to export all patients in .3sz in Ortho System 2021.1

Published November 7th, 2023 by Filip Poswiata

Export patient information in Ortho system 2021.1 Start Ortho System 2021.1, select 'Patients' tab, right-click on any patient, and select 'Export All' and 'Export to .3sz' file. You will be directed to the 'Export' window where all patients and model sets will be selected for export by default. It is possible to customize the number of patients/mod

5808 Views 1 min

What folders are shared during Dental System Server installation

Published November 8th, 2023 by Filip Poswiata

Sharing and Setting Permissions for Dental System Components on Multiple PCs If Dental System components are installed on different PCs, the component folders should be shared and available with Full Control for Everyone. When performing a Server installation, the Dental System folders are shared automatically.  The following folders will be automat

10361 Views 1 min

"TRIOSScanData: Stream read error." when exporting any TRIOS scan in Dental Manager

Published November 8th, 2023 by Filip Poswiata

Description of the error  This error is shown when trying to export Trios scans on a disk with no space left. It may also appear for other export options such as CAD Output   Solution: Export file on a disk with enough space available or free some space on the desired disk.

4567 Views 1 min

"Access Violation in module GenericModule.dll"

Published November 8th, 2023 by Filip Poswiata

"Access Violation in module GenericModule.dll" The ‘Access Violation in GenericModule.dll’ error randomly happens in Dental System when opening an order in DentalDesigner and other DS applications, or trying to generate CAM output, or when closing DS applications. The root cause for this is a known bug. Affected DS versions: DS 2017 (all patches), a

5515 Views 1 min

"Unable to create a screenshot. Please ensure that you use the latest video/OpenGL drivers." in Dental Designer

Published November 8th, 2023 by Filip Poswiata

Root Cause: Error message "Unable to create a screenshot. Please ensure that you use the latest video/OpenGL drivers." arises due to the use of a non-NVIDIA graphics card or an outdated graphics card driver.  Affected Versions: All versions.  Possible Solutions: Graphics performance settings  Please ensure that graphic performance is set to “High” f

5915 Views 1 min

Error Message: "Invalid image size" when opening any model set in Ortho Analyzer

Published November 7th, 2023 by Filip Poswiata

Understanding image size in Ortho Analyzer The error occurs when the PC screen resolution setting is not set to 100%, to resolve the issue: Click on the Windows button to bring up start menu. Click on Appearance and Personalization. Under Display, click on Make text and other items larger or smaller. Adjust the magnification settings of Smaller (100

4580 Views 1 min

How to export prepared model in Ortho System

Published November 7th, 2023 by Filip Poswiata

After preparing the virtual base model, you have two options for exporting it as an STL file: By exporting immediately via the export option within the software when you finish the workflow. By saving your work and exporting the model at a later time. Exporting via the software after finishing the virtual base workflow Step 1: Click the export model

5770 Views 1 min

Access Violation or Gingiva line violation when trying to create a gingiva base for full dentures

Published November 8th, 2023 by Filip Poswiata

When trying to create a gingiva base for the full dentures at the FD Initial Setup step a warning appears: "Gingiva lines violation" + EM: “Access Violation at the address”.     This issue is caused by a lack of spacing between the teeth and the gingiva as seen in this picture. The gap is needed when designing Full Dentures. Solution Step 1: In the

5691 Views 1 min

"Exception TPen in module ModelBuilder.exe" in ModelBuilder

Published November 8th, 2023 by Filip Poswiata

"Exception TPen in module ModelBuilder.exe" in ModelBuilder In some cases there can be an error message in ModelBuilder: "Exception TPen in module ModelBuilder.exe", mostly on Preparing Model step: Affected DS versions: any version. Potential root causes: bad scan quality, holes/artifacts in scans, bad trimming of scans. Possible solutions: Smooth o

5970 Views 1 min

Unable to load XML document (XML document must have a top element) In Dental Manager when trying to design an order

Published November 8th, 2023 by Filip Poswiata

Issues with the design step of a scanned order The following error message may show up when trying to proceed the design step of a scanned order in Dental Manager: Root cause: Corrupted scans, corrupted XML files in the order. Affected Dental System versions: Potentially all Dental System versions. Possible solutions: Ask the customer to create the

6190 Views 1 min

Order sent to manufacturer in DS-2018 or later does not show up on the FTP server

Published November 8th, 2023 by Filip Poswiata

Description of the issue In Dental System 2018 or later the user is pressing "Send to Manufacturer" button for the designed order, and after some time the software informs that sending was completed successfully. However, the order is not visible on the FTP server of the manufacturer. If both 3Shape Communicate and FTP Inbox credentials are provided

4707 Views 1 min

How do Copy options in Dental Manager work

Published November 8th, 2023 by Filip Poswiata

There are several options of copying an order in Dental Manager. All of them can be accessed by right-clicking on the order and going to Copy:  Standard copy This option will create a copy of the order with all the files and settings of the original order. If the original order was in the 'Designed' state its copy will be in 'Scanned' state, but as

9162 Views 1 min

"Current dongle configuration doesn't allow to save the design"

Published November 8th, 2023 by Filip Poswiata

"Current dongle configuration doesn't allow to save the design" When trying to design Abutments at the beginning of design pops up the EM: "Current dongle configuration doesn't allow to save the design". Possible reasons: The dongle is either missing or there are no more license available in the dongle. Your current license setup does not support th

5033 Views 1 min

During the design in Dental Designer - part of the model and design disappear.

Published November 8th, 2023 by Filip Poswiata

Scan and designed teeth disappear at the Anatomy Design At the Anatomy Design step the part of scan and designed teeth disappear. Solution Go to Anatomy Design step and disable the scissors from the Visualization Slider - When it is active, it cut the neighbors of the model, so you can see it from the sides when rotating: More information can be fou

5180 Views 1 min

Error: The part is signed with "[Provider name]" and the corresponding option is not enabled on the site or the local key file is not updated

Published November 8th, 2023 by Filip Poswiata

This error can pop up in Dental Designer or in the Dental System Control Panel (when checking a specific Implant System). It appears because your dongle is not configured (has no permissions) to use a specific Implant library of the Provider mentioned in the error message. Solution Contact the library provider mentioned in the error message and ask

5614 Views 1 min

"Program version not enabled in dongle, please downgrade" when opening Dental System

Published November 8th, 2023 by Filip Poswiata

"Program version not enabled in dongle, please downgrade" when opening Dental System Error message "Program version not enabled in dongle, please downgrade" when trying to open Dental System. Root cause: The currently installed DS version is not allowed on the connected dongle (there is a version limitation enabled on the dongle).  Solution: contact

4979 Views 1 min

Empty RPD orders arrive with an error "Incomplete/Invalid, Open(Ctrl+M) to correct, Model element is not connected" in 3Shape Communicate Inbox

Published November 8th, 2023 by Filip Poswiata

Empty RPD orders arrive with an error "Incomplete/Invalid, Open(Ctrl+M) to correct, Model element is not connected" in 3Shape Communicate Inbox Some of the RPD orders arrive in 3Shape Communicate Inbox as empty even if it is scanned or designed by sender. Affected Dental System Version: Potentially any version Root Cause: RPD Frame element is not pr

4703 Views 1 min

"Incomplete or invalid order" in Dental Manager TRIOS Inbox arriving from Implant System

Published November 8th, 2023 by Filip Poswiata

Description of the issue Error Message: "Incomplete or invalid order. Message code 30:73. Title: - AbutmentKitID for the abutment is empty. Description: AbutmentKitID for the abutment on tooth "x" is empty. Additional message: Not possible to add an abutment on tooth "x", because the implant system with Manufacturer="y", System="z" and PartType="a"

4712 Views 1 min

"Stream read error" when opening Dental Manager or Dental System Control Panel

Published November 8th, 2023 by Filip Poswiata

Description of the issue When trying to open Dental Manager or Control Panel below error is shown and software is closed. Root of issue: No free space on a local disk (it might be a different disk where Dental System is located) Affected versions: Potentially all Solution: Free some space on the full disk

5345 Views 1 min

"Access violation in module DentalDesigner.exe" in Dental Designer (general article)

Published November 8th, 2023 by Filip Poswiata

Dealing with access violation "Access violation at address XXXXXXXXXXXXXX in module "DentalDesigner.exe". Read of address XXXXXXXXXXXXXXX" may be popping up at any stage of the design workflow. This is a very generic error message indicating that something went wrong during the design process of the specific order, and may not be always reproducible

8861 Views 1 min

How do user manuals in Dental System 2019 and above work

Published November 8th, 2023 by Filip Poswiata

Accessing Technical Documentation and User Manuals in DS-2019 and above In DS-2019 there has been a change to the user manuals and now there are 2 documents available when pressing the F1 key: Technical documentation (Lab Solution) - this is a short version with CE mark: Available in the following languages: English, Chinese, French, German, Greek,

5761 Views 1 min

Duet Display prevents Dental System applications from opening

Published November 8th, 2023 by Filip Poswiata

Duet Display prevents Dental System applications from opening When having the Duet Display program installed on a PC with Dental System, some of the applications like Dental Designer, Model Builder or Splint Designer cannot be open. There is no error message, expert for a pop-up saying that the application "didn't make any output files": Affected DS

4865 Views 1 min

"An error occurred while CAM output generation. Error message: The specified path was not found" when generating CAM output

Published November 8th, 2023 by Filip Poswiata

Issue with generating CAM output / F7 When in Dental Manager, trying to generate CAM output / F7, error message is shown:  Manufacturing output could not be generated.  RawSTL: An error ocurred while CAM output generation.  Error message: The Specified path was not found Root cause: The path is not configured correctly in Dental system control panel

5701 Views 1 min

EM "Order not supported. Invalid boundary (Boundary.Count=2)" when opening an order in Model Builder

Published November 8th, 2023 by Filip Poswiata

EM "Order not supported. Invalid boundary (Boundary.Count=2)"  or "Order non supported. Invalid boundary (Boundary.Count=1)"  happens when opening Model Builder after designing restorations in Dental Designer (no matter which restoration). When opening Model Builder with this Error Message, model design is not possible.  Root cause: Poor scan qualit

7436 Views 1 min

Missing scans in a TRIOS order coming to 3Shape Communicate Inbox

Published November 8th, 2023 by Filip Poswiata

Missing scans in a TRIOS order coming to 3Shape Communicate Inbox When an order from TRIOS comes in to dental manager communicate inbox showing the error below: Affected DS versions: potentially all DS versions. Scenario 1: All scans can be seen in the 3scom portal.  check the order in the 3scom portal to know if the all the scans are included. If n

6208 Views 1 min

How to prepare Virtual Model Bases in Ortho Analyzer

Published November 7th, 2023 by Filip Poswiata

Create a virtual base in the Ortho System In order to create the virtual base please right-click on the model set and select the  prepare model set  option.  The first step in order to prepare the base will be the plane alignment for both the occlusion and sagittal plane:  Step 1: In order to select the occlusion plane three points need to be select

7353 Views 1 min

Orders randomly not arriving into 3Shape Communicate Inbox

Published November 8th, 2023 by Filip Poswiata

Orders randomly not arriving into 3Shape Communicate Inbox There can be several random orders that are not arriving into 3Shape Communicate Inbox although they are visible on 3Shape Communicate web portal.  This article describes only randomly missing orders, while there can be various reasons for orders not arriving at all (e.g. missing internet co

6399 Views 1 min

Hollowing models in Model Builder

Published November 8th, 2023 by Filip Poswiata

Hollowing models in Model Builder In case it is required to hollow the models, such option is available in Model Builder. Usually it is done to minimize the quantity of the used material. Such option can be found in the left upper corner in File > Virtual Trimming preferences > Model Manufacture> Hollowing > change Surface thickness valu

10356 Views 1 min

Unable to select a specific appliance type in Appliance Designer

Published November 7th, 2023 by Filip Poswiata

If a specific appliance type does not appear in the drop-down list when selecting a new appliance (see image below), it means the required appliance type has not been downloaded in the Ortho Control Panel. This issue might also occur with an Archwire or a bracket library. Verification Before resolving the issue you must first verify if this is the c

4321 Views 1 min

How to import STL files as attachments for Ortho System

Published November 7th, 2023 by Filip Poswiata

Import STL files as attachments in the Ortho System This article will guide you through the steps for the import of STL files as attachments in the Ortho System. Steps to be taken: The customer/reseller must have the file in the STL format in the Attachments folder located in the following path: C:\ProgramData\3Shape\Ortho Control Panel\library\3Sha

7903 Views 1 min

"The manufacturing process for the order is not in the list of allowed processes" when starting to design

Published November 8th, 2023 by Filip Poswiata

Unable to load file  When opening a scanned order in Dental Designer, there may be an error message "Unable to load file (The manufacturing process for the order is not in the list of the allowed processes)". Affected DS version: any version. Selected manufacturing type: local. Root cause: manufacturing process method (CADOutput .dll file) for the c

5992 Views 1 min

"Error running Splint Studio. Path to Splint Studio was not set" when trying to start design in Splint Studio from DS

Published November 8th, 2023 by Filip Poswiata

"Error running Splint Studio. Path to Splint Studio was not set" when trying to start design in Splint Studio from Dental System Dental System 2.19.3.0 and higher has a full integration with Splint Studio based on 3Shape Dental Desktop - i.e. the splint order can be created and scanned in Dental System and then designed in Splint Studio. ! NOTE: bef

5022 Views 1 min

Populating Anatomy and Pontic Library

Published November 8th, 2023 by Filip Poswiata

Populating Anatomy and Pontic Library To populate an empty Anatomy and Pontic library or modify one, you can either scan a model or import a scanned model. Model Scanning: To scan a model: Open ScanIt Library by selecting Anatomy elements->ScanIt library in Dental System Control Panel,click the Scan crown library button. OR In Dental System Contr

7926 Views 1 min

Clear Aligner Studio Workflow in Ortho System

Published November 7th, 2023 by Filip Poswiata

Explore clear aligner opportunities  Clear aligners represent one of the fastest growing areas within dentistry. And with the recent expired patents, new opportunities for labs and clinics have been opened to enter this highly profitable market.  3Shape's all-in-one solution 3Shape introduced the innovative and powerful all in-one software solution

6988 Views 1 min

Opening the software on multiple computers in a server-client setup

Published November 7th, 2023 by Filip Poswiata

Simultaneous software access: additional seats In order to be able to open the software simultaneously on multiple computers a client point of the module is required for each PC where the software will be opened. For example, to open the software on 3 PC's simultaneously on a single server with two clients, 3 client points would be required. The sof

5548 Views 1 min

Order can't be accepted automatically in 3Shape Communicate Inbox

Published November 8th, 2023 by Filip Poswiata

Orders can't be accepted automatically in 3Shape Communicate Inbox Some order arriving in 3Shape Communicate Inbox may not be accepted automatically despite the "Inbox Auto-accept" option is enabled in Dental System Control Panel > Tools > 3Shape Communicate.  The order will remain in the Inbox and there can be an error message: "Order require

5467 Views 1 min

"Invalid boundary" after placing cut out line in Dental Designer

Published November 8th, 2023 by Filip Poswiata

Invalid boundary error "Invalid boundary" error appears when trying to proceed to the next step after placing cut out line. Frequently happening on temporaries design, where the teeth have to be removed from the scan. Affected DS version: potentially any version. Possible root cause: there is a gap between the teeth in the scan and the cut out lines

7944 Views 1 min

"Retrieving orders failed" in Dental System Communicate Inbox

Published November 8th, 2023 by Filip Poswiata

"Retrieving orders failed" in Dental System Communicate Inbox New orders do not arrive from the Portal and the message "Retrieving orders failed" is displayed in the upper-right corner of Dental System Communicate Inbox. Affected DS versions: potentially all versions. Possible root cause: Some orders in Communicate Inbox directory are not valid for

6225 Views 1 min

"Incomplete or invalid order" in Communicate inbox in Dental Manager, that is sent from Implant Studio

Published November 8th, 2023 by Filip Poswiata

Description of the issue  "Incomplete or invalid order. Colour has not been set for the following elements ( which is mandatory for their materials " in Communicate inbox in Dental Manager, sent from Implant Studio. Potential root cause:  Only surgical guide and implant are chosen in the order form in Implant Studio. Dental System is looking for abu

4633 Views 1 min

Troubleshooting "No Subscription" Error

Published November 8th, 2023 by Filip Poswiata

Troubleshooting "No Subscription" Error When opening Dental Manager or Dental System Control Panel there is an error: "No subscription for this application the dongle is either missing or there are no more application subscription available in the dongle. Please set the 3Shape Dongle Service under System Settings. [Socket Error #11001 Host not found

6693 Views 1 min

Invalid Mesh Error During the Validation of Custom Attachments

Published November 7th, 2023 by Filip Poswiata

Troubleshooting invalid mesh issue in Ortho Control Panel During the validation of custom attachments in Ortho Control Panel sometimes the error message “Model has invalid mesh (BadConfiguration)/(SelfIntersections)” will appear. This happens if the software detects an error in the imported SLT files, resulting in the file not being imported.  To fi

4204 Views 1 min

"Cannot generate the CAM output for [indication] because [library name] is not allowed to be produced locally" when generating CAM output

Published November 8th, 2023 by Filip Poswiata

"Cannot generate the CAM output for [indication] because [library name] is not allowed to be produced locally" when generating CAM output This kind of warning can appear and prevent the user to generate CAM output after an order has been designed: Affected DS versions: any version. Selected manufacturing type: local. Root cause:  selected library ha

4881 Views 1 min

How to change the sub-setups animation length

Published November 7th, 2023 by Filip Poswiata

This article applies to Ortho system versions 2019 and older.   The Ortho system includes a Sub-Setups Animation tool that visualizes treatment planning in virtual setups. By default, the animation runs for 10 seconds, regardless of how many sub-setups are included in the treatment plan. You can adjust the animation length in the Ortho Control Panel

5007 Views 1 min

How to backup Ortho System files and settings

Published November 7th, 2023 by Filip Poswiata

This article applies to OS versions older than 2021.   Backup Ortho Data folder The Ortho Data folder is a physical storage of patient’s data. By default it is installed under the: C:\ProgramData\3Shape\OrthoData\ Ortho Data may have different folder names, and different locations, and can be pointed to another PC within the same network. Always che

7328 Views 1 min

Secondary 3Shape Communicate account in Dental System

Published November 8th, 2023 by Filip Poswiata

Secondary 3Shape Communicate account in Dental System Dental System 2021 introduces the possibility to register multiple 3Shape Communicate accounts and connect them to the same Dental Manager Inbox. This feature will make it easier to exchange orders using different TRIOS order templates (for example, for customers from different regions) with diff

5592 Views 1 min

Where can I find an updated list of System Requirements for Ortho System?

Published November 7th, 2023 by Filip Poswiata

System requirements can be found in the User Manual (start OA or AD 2017-1). To access: Right click on 3Shape logo in the bottom right corner Select User Manual And follow the path: Home > System Settings > System Requirements.

6460 Views 1 min

"Connection Closed Gracefully" or "Out of memory while expanding memory system" with Avira antivirus installed

Published November 8th, 2023 by Filip Poswiata

"Connection Closed Gracefully" or "Out of memory while expanding memory system" with Avira antivirus installed   Avira Web Protection option of Avira antivirus may block some of the 3Shape Dental System modules, such as Dental Manager, Dental System Control Panel, Dental Desktop, ScanItRestoration, etc.  There is no way to configure Avira so that it

5055 Views 1 min

'Cannot upload file to remote site xxxxx' Error message when trying to send an order through FTP from lab to lab

Published November 8th, 2023 by Filip Poswiata

Error with sending orders When trying to send an order from one lab's dental manager to another using FTP, sometimes the error "Unable to upload file to remote site xxxx" will show up. Solution Open Dental System Control Panel. Look for the external lab/manufacturer which is giving the error. Disable the 3Shape communicate integration for the extern

5527 Views 1 min

E-scanners and Ortho System Compatibility

Published November 7th, 2023 by Filip Poswiata

E1,E2,E3 compatibility E1, E2, E3 are supported only with Ortho System 2015-1 patch 9 or higher. They are not compatible with older versions of Ortho System as well as Scan Suite older than 10.1.0. Regarding the compatibility with other software: E scanners + Scan Suite 10.1.0 are supported only with Dental System 2.16.4.0 and higher. E4-series comp

806 Views 1 min

"Communication failed. HTTP/1.1 403 Forbidden" when saving 3Shape Communicate credentials in DSCP

Published November 8th, 2023 by Filip Poswiata

"Communication failed. HTTP/1.1 403 Forbidden" when saving 3Shape Communicate credentials in DSCP When saving 3Shape Communicate credentials in Dental System Control Panel > 3Shape Communicate, the password containing some symbols (e.g., % or #) is not accepted and the following error message appears: "Communication failed. HTTP/1.1 403 Forbidden

4423 Views 1 min
Load More